Document Type

Article

Department

Pathology and Microbiology; Pathology and Laboratory Medicine

Abstract

Point-of-care testing (POCT) coagulometers are increasingly being used in the hospital setting and patients’ self-testing. We determined the agreement of prothrombin time international normalized ratio (INR) results by POCT coagulometer and laboratory instrument through a comparative analysis and investigated whether the results of POCT coagulometer can reliably be used without being confirmed by standard laboratory analyzer. A total of 200 INR measurements by POCT coagulometer (CoaguChek XS Pro) and laboratory analyzer (Sysmex CS2000i) were compared using Passing-Bablok regression analysis and Bland-Altman plot. Agreement of the INR measurement was further analyzed in relation to dosing decision. The correlation of INR measurements between CoaguChek XS Pro and Sysmex CS2000i was excellent (correlation coefficient ¼ 0.973). The overall mean difference was 0.21 INR + 0.32 (range: 1.7-0.44). The mean difference was found to get increased as INR results increased and was 0.09 in the subtherapeutic range (1.9 INR), 0.29 INR in the therapeutic range (2.0-3.0 INR), while 0.4 INR in the supratherapeutic range (>3.0 INR). The overall agreement was excellent (k ¼ 0.916) and overall 11 (5.5%) of 200 INR measurements showed a difference in dosing decision between the 2 instruments. The positive bias of POC-INR is evident in the supratherapeutic range which could affect the dosing decision requiring confirmation with the laboratory INR measurement.